The Spectra-Physics® Inspire™ ultrafast OPO delivers user-friendly, gap-free computer-controlled tuning from 345 nm to 2.5 μm with no intracavity optics or crystal change. Pumped with the Mai Tai® HP Ti:Sapphire laser, the OPO delivers high power across the UV and visible range with adjustable pulse widths from 80 to 250 fs. Inspire’s robust opto-mechanical design ensures high environmental stability.
Inspire Auto 50 | Inspire Auto 100 | Inspire HF 50 | Inspire HF 100 | |
---|---|---|---|---|
Average Power | ||||
SHG @ 400 nm | N/A | 1100 mW | N/A | 1100 mW |
Signal @ 550 nm | 350 mW | 350 mW | 350 mW | 350 mW |
Depleted Fundamental @ 800 nm | 1100 mW | 1100 mW | 1100 mW | 1100 mW |
Idler (at peak of curve) | 170 mW | 170 mW | 170 mW | 170 mW |
Pulse Width | ||||
SHG | N/A | <140 fs | N/A | <140 fs |
Signal | 100-250 fs (adjustable) | 100-250 fs (adjustable) | 200 fs | 200 fs |
Depleted Fundamental | <140 fs | <140 fs | <140 fs | <140 fs |
Idler | 80-250 fs (adjustable) | 80-250 fs (adjustable) | 200 fs | 200 fs |
Tuning Range | ||||
SHG | N/A | 345-520 nm | N/A | 345-520 nm |
Signal (Simultaneous with Idler) | 490-750 nm | 490-750 nm | 490-750 nm | 490-750 nm |
Depleted Fundamental | 690-1040 nm | 690-1040 nm | 690-1040 nm | 690-1040 nm |
Idler (Simultaneous with Signal) | 930-2500 nm | 930-2500 nm | 930-2500 nm | 930-2500 nm |
Repetition Rate | 80 MHz | |||
Noise | <1% rms | |||
Wavelength Stability @ 555 nm | <0.5 nm | |||
Spatial Mode | TEMoo, M2 <1.2 (Signal) | |||
Polarization | Horizontal for Signal and Idler Vertical for SHG |
|||
Spectrometer for UV and Visible Range3 | 350-900 nm (integrated into optics unit) | |||
Dimensions (W x L x H)4 | 14.2 x 37.6 x 8.1 in (36.0 x 95.4 x 20.7 cm) |
Footnotes:
To verify and analyze the output power of your Inspire UV laser, we recommend the Ophir 3A (7Z02621) thermal power sensor and Centauri laser power meter. Additional information on this laser measurement equipment can be found on the ophiropt.com website. Other Ophir sensors and meters may also be suitable for the Inspire UV, depending on application. Ophir also works with laser system developers to achieve embedded beam diagnostics to monitor system performance as an OEM solution.