Applications

Specifications

Output Characteristics

HighQ-2-IR HighQ-2-SHG
Average Power >1.5 W >0.65 W
Wavelength 1045 ±8.0 nm 522 ±4.0 nm
Repetition Rate 63 MHz
Pulse Width (FWHM) <250 fs
Pulse Energy >24 nJ >10 nJ
Peak Power >80 kW >35 kW
Power Stability <1% rms (100 hours)
<0.5% rms (12 hours)
Beam Quality TEM00, M2 <1.15
Beam Diameter, at waist 0.8 mm 1.0 mm
Polarization 100:1, vertical3 100:1, horizontal
Ellipticity <10%
Cold Start Time ≤20 min
Warm Start Time ≤15 min
Operating Temperature Range 18-32°C

Cooling Requirements

HighQ-2-IR HighQ-2-SHG
Laser Head Air-cooled
Power Supply Air-cooled

Utility Requirements

HighQ-2-IR HighQ-2-SHG
Voltage 100-240 VAC, 50/60 Hz, 175 W

Laser Head

HighQ-2-IR HighQ-2-SHG
Dimension (L x W x H) 8.74 x 8.03 x 3.97 in (222 x 204 x 100.9 mm)
Weight 13.7 lbs (6.2 kg)

Power Supply Physical Characteristics

HighQ-2-IR HighQ-2-SHG
Dimension (L x W x H) 19 in rack mountable, 2U.
18.03 x 19.02 x 3.46 in (458 x 483 x 88 mm)
Weight 28.2 lbs (12.8 kg)

Footnotes:

  1. Due to our continuous improvement program, specifications are subject to change without notice.
  2. HighQ-2 is a Class IV - High Power Laser, whose beam is, by definition, a safety and fire hazard. Take precautions to prevent exposure to direct and reflected beams. Diffuse as well as specular reflections can cause severe skin or eye damage.

Features

24/7 Operation

Designed as a reliable OEM laser for 24/7 operation, the HighQ-2 laser combines true turn-key operation and fast cold start turn-on, ensuring quick installation and easy operation. HighQ-2 is equipped with long life diodes, a sealed optical cavity design and is manufactured in a controlled, cleanroom environment. The result  is a highly dependable laser with long lifetime, high uptime and low cost of ownership.

Compatible Power Sensors

To verify and analyze the output power of your HighQ-2 laser, we recommend the Ophir 3A-PF-12 (7Z02720) thermal power sensor and Centauri laser power meter.  Additional information on this laser measurement equipment can be found on the ophiropt.com website.  Other Ophir sensors and meters may also be suitable for the HighQ-2, depending on application. Ophir also works with laser system developers to achieve embedded beam diagnostics to monitor system performance as an OEM solution.